4 configuring the alarm, 5 offset calibration – HF scientific AccUView Wastewater UV %Transmission Monitor User Manual

Page 25

Advertising
background image







To enable the Modbus mode, select

ASCII or RTU. For more information refer to the

Modbus Manual (Catalog #24570). The manual is also available as a free download from
our website at

www.hfscientific.com

.


7.4 Configuring the Alarm
A relay is provided and is designed to operate as an independent programmable alarm.
Two types of information must be input to fully program the alarm:

1. The alarm function (HI, LO, or OFF)
2. The alarm set point (level at which the alarm activates)
These are described below:

Alarm Function: The alarm can either be turned OFF or programmed to operate in one
of two different manners:

1. HI alarm: the relay changes state when the measured %T level is higher than the

programmed alarm level for a prescribed amount of time.

2. LO alarm: the relay changes state when the measured %T level is lower than the

programmed alarm level for a prescribed amount of time.

Note: The relay automatically changes state when an internal system failure is detected.

Alarm Set Point: The level at which an alarm activates is called the alarm set point. On
the instrument, the alarm set point is designated as “S/P”. The set point is adjustable to
any valid level over the range of the instrument in steps of 0.1%T.

7.5 Offset Calibration
Refer to section 6.0 for more information on this selection.










AccUView Wastewater (1/09)

19

Rev. 2.9