MODEL 110E PUMP OPERATION / MAINTENANCE MANUAL

Figure 1-1

ATTENTION: The pump should be operated with clean, dry air or nitrogen.
Particulate, water and oils in the air supply can damage the pump.

NOTE:

1. It is recommended that a filter be placed on the discharge side of the pump.

2. Although extensive efforts are made to deliver pumps to our customers
completely dry, new pumps may contain residual moisture from their final DI
water test.

Recommended Maximum Operating Levels:

Maximum supply CDA/N2 pressure: 60 psig (4.1 bar)

Maximum fluid temperature: 212°F (100°C)

See Section 4.1 Preventative Maintenance Schedule for additional requirements
for high temperature applications.