1 calibrating linear voltage inputs, 1 set channel, 2 zero cal – Monarch Instrument DC6000 User Manual

Page 112: 3 full cal, Calibrating linear voltage inputs, Set channel, Zero cal, Full cal

Advertising
background image

Section 7 - Instrument Care and Maintenance

I.

Page 104

1/06/2012 REV 1.10.2

1 Hz Cal

– Calibrate the units frequency reference. Apply a 1 Hz signal to channel 1, 6, 7 or 12. Doing this

on one channel calibrates all four channels.

Store Cal

– This saves the new calibration settings into non-volatile memory.

Set Channel:

Activate the Channel to be calibrated. This will enable the Live Data.

Calibrate: Perform the calibration

– This is non-reversible.

Live Data: Shows the actual live input on the selected Channel.

Restore Factory Calibration:

Will restore the calibration data that was in effect when the unit shipped

from the factory

by pressing the “Restore” button

Note: Restoring the factory calibration values may negate the current calibration of your unit unless the
factory calibrates the unit on a regular basis.

7.3.1 Calibrating Linear Voltage Inputs:

You will need a shorting jumper (no connection to COM) and the voltage calibrator.

7.3.1.1

Set Channel

Select the Channel the calibrator is connected to and
select the Voltage Range to be calibrated from the
drop down menus

and press the “Set Chan” button.

The “Live Data” window will show the
current value from the connected
source and a “Channel has been set
successfully” popup will appear.
Press “OK” in the top right corner

7.3.1.2

Zero Cal

Select the Zero Cal function from the Calibration
Function drop down menu.

Apply 0 Volts (Shorting jumper or
set Calibrator to 0V) to the channel
selected in step one and press the
“Calibrate” button. Doing this on
one channel calibrates all channels. On completion of the step you will get a

“Calibration Function OK" popup dialog. Press “OK” in the top right corner.

7.3.1.3

Full Cal

Select the Full Cal function from the Calibration Function drop down menu.

Select the Range to be calibrated from the Range drop down menu.

Apply the Full Scale Input voltage for the selected range from the table
to the right. The output of the calibrator is connected to the IN+ and
IN- inputs on the selected channel. There is NO CONNECTION to
COM. (See

Wiring Section

for connection detail). Note that the full

scale calibration value is greater than the voltage range.

The “Live

Data” window will show the current value from the connected source.
P

ress the “Calibrate” button to complete the calibration. Doing this on

one channel calibrates all channels.

Accept the “Calibration Function

OK” popup dialog.

To calibrate another range repeat from step 6.3.1.3. Calibrate all or as
many ranges as you need.

Voltage Range Full Scale Input

±125mV

146mV

±250mV

292.5mV

±500mV

585mV

±1.00V

1.170V

±3.00V

3.125V

±6.00V

6.250V

±12.00V

12.50V

±24.00V

25.00V

Figure 7.2

– Calibration Window