Appendix a – technical specifications, Display, Power – Monarch Instrument DC6000 User Manual

Page 115: Analog inputs

Advertising
background image

Appendix A - Technical Specifications

I.

Page 107

1/06/2012 REV 1.10.2

APPENDIX A – TECHNICAL SPECIFICATIONS

DISPLAY

Display Type Color

CCFL backlit Active Matrix TFT Liquid Crystal Display (5.6 inch) QVGA with resistive
analog touch screen control.

Resolution Color

320 (H) X 234 (V) pixels.

Display Modes

Graphics (Trending vertical or horizontal), Bar Graphs (vertical or horizontal, large or
small), Digital Meter (large or small), Alphanumeric Alarm and Event Log. Display builder
allows user to create custom displays

Virtual Chart Speed

Programmable: 0.5 in/hr to 600 in/hr (10 mm/hr to 15,000 mm/hr) ??

Display Windows

Time/Date, Graphics (Bars, Large Digital, Trends), Disk Status, System Status, Menu
Button Bar, Unit Identification, Alarms/Events

POWER

Power Requirements 100 to 240 Vac, 50/60 Hz. 35 VA max. Optional 24 Vdc + 15%

Power Fail Protection Programmed parameters stored in non-volatile memory. Clock battery backed. Internal

battery backup provides orderly shutdown and the ability to survive brownouts and short
blackouts (<20 seconds). Data retention time without power 3 months

Power Output

Optional isolated 24Vdc @100mA output

ANALOG INPUTS

Number of Inputs

6 or 12 direct inputs plus 6 additional calculated channels

Measurement Category: CAT I

Maximum Input

30 Vdc. Transient OverVoltage Protection - 50Vdc

Isolation

Installation category II: Pollution Degree 2
350 Vdc or RMS AC channel to channel, 1500 Vdc or peak AC Channel to Chassis

Input Resolution

0.0015% of full scale, 16 bit unless otherwise stated

Input Impedance

>1megohm

Measurement Rate

10 times per second on all direct input channels

Common Mode Noise Rejection

>100 dB, 50/60 Hz filter enabled

Normal Mode Noise Rejection

>50 dB at 50/60 Hz, filter enabled

Math Functions

fully programmable +, -, x, /, square root, sine, cosine, tangent, log, totalization, powers,
averages, conditional logic

– AND, NOT, OR, +, >, <, gated timers. Can use live channels

in calculation. Can define 6 constants and 6 functions per channel.

Analog Input Signals: DC Voltage:

Bipolar Input

Full scale ranges: +125mV, +250mV, +500mV +1.00V, +3.0V, +6.0V, +12.0V, +24.0V

Accuracy:

Ranges to 1V ±0.06%:
Ranges > 1V ±0.1%

DC Current:

(with external shunt) 4 to 20 mA, 0 to 20 mA and 10 to 50 mA
Accuracy ±0.15% using external 50 ohm 0.1% shunt