Understanding the keys on the control unit, Calibrating the sensor – ADS Environmental Services FlowShark HV QR 775005 A2 User Manual

Page 16

Advertising
background image

16

ADS FlowShark HV O&M Manual

Understanding the Keys on the Control Unit

The following table provides general functional descriptions for the keys on the
control unit. Each key serves multiple functions. The current function of a key
depends on the current content displayed on the control unit screen.

Key

Functional Description

ESC

Turns off the control unit when held down for more than 2 seconds

Returns the menu to the previous level

Turns on the control unit when held down for at least 5 seconds

Displays the parameter settings from the top level

Displays subsettings of parameter settings (when applicable)

Displays current value/option setting corresponding to parameter setting selected

Activates editing mode for bracketed text/values ([xxx])

Deactivates editing mode for bracketed text/values (<xxx>) and saves text/values for
parameter

Scrolls forward through top level screens

Scrolls forward through available parameter settings and subsettings

Moves cursor forward through individual characters in editable field

Scrolls backward through available parameter settings and subsettings

Moves cursor backward through individual characters in editable field

Scrolls forward through individual characters or options in editable field

Scrolls forward through available options for selection in an editable field representing a
particular parameter setting

Scrolls backward through individual characters or options in editable field

Scrolls backward through available options for selection in an editable field representing a
particular parameter setting

Calibrating the Sensor

Before taking velocity measurements, immerse the sensor into the flow to acclimate
the FlowShark HV’s integrated temperature sensor to the temperature of the flow.
Temperature differences between the sensor and flow can produce inaccurate velocity
measurements. A difference of only 1°C can lead to a measurement error of 0.24%.

Acclimation is complete when the temperature reading displayed begins to stabilize
and deviates only within a very limited temperature range of a few degrees.
Temperature and velocity stabilization may require 1 to 5 minutes upon initial use. It
can require several minutes if a large difference exists between the temperature of the
flow and the temperature at the location outside the flow.