Oshpd anchor pre-approval opa-1743 – Amico AHM Monitor Arms (Adjustable Height) User Manual

Page 7

Advertising
background image

www.amico.com

4

Section 2: Installation on Rail System

Mounting to the Monitor Rail System (MRS)

with the MON-Adapter

WARNING:

Never attempt to mount or remove the arm from

the rail when a monitor is installed.

WARNING:

To prevent the adapter from falling down the

channel, ensure that the 4 set screws are fastened to the
tightest possible position

(Figure 1)

.

NOTE:

For safety, top and bottom stoppers prevent the adapter

from sliding out of the MRS. To remove, the adapter must be
taken out from the top of the rail with the top stopper pressed
back into the rail

(Figure 2)

.

1. With the AHM arm locked in the highest vertical position,

support the bottom of the arm and guide the MON-channel

adapter into the top of the wall channel.

2. Once the arm is at the desired height, tighten all four screws

on the MON-channel adapter using a 1/8” (3 mm) hex key

(Figure 1).

OSHPD Anchor Pre-Approval OPA-1743

In compliance with standards established for the anchor and installation of instrument support systems by the
California Office of Statewide Health Planning and Development (OSHPD), Amico Accessories Inc. has obtained pre-
approval for MRS 19’’ Wall Channels or longer with extension Arm. For proper installation requirements please refer to
our OPA Document.

(Figure 2)

Top stopper

Bottom stopper

MRS Rail: Length for representational use only

(Figure 1)

Standard MON-Adapter

MRS Rail

Set Screws (3
mm)