Amico Condor Adjustable Height Channel User Manual

Page 17

Advertising
background image

Preventative Maintenance

WARNING

The Condor requires periodic inspection and maintenance to perform optimally and achieve

maximum operation life. The intervals shown below are recommended. Maintenance schedules

should be shortened for Condors that see heavy use.

NOTE:

Please follow the Preventative maintenance protocol for the accessories as stated in the manual for those

accessories.

Please be sure to thoroughly check the areas illustrated below for Condors.

Area 1: Mounting Channel and Track System

a. Visually inspect for any signs of grinding and gapping along the sides of the channel.
b. Check if the accessories are mounted properly and tightly to the channel.
c. Remove the MRS cover and ensure that the cables are staying inside the tracks and are not tangled.

Area 2: Cable Management Clamps

a. Visually inspect for any signs of grinding by checking for debris around the part.
b. Make sure the screws are tightly installed and the cables are clamped down properly.

Area 3: Cable Management Side

a. Remove cover and check if there is grinding or tangling of cables inside.

Area 4: Force Adjuster Side

a. Visually inspect for any signs of grinding and gapping on the pulleys as well as on the side of extrusions.
b.

Pulleys:

Ensure the screws holding the pulleys are fastened tightly.

c.

Spring:

Visually inspect the spring to see if there has been any damage to the outer shell.

d.

Threaded rod:

check for bending and damage to the threads.

SECTION 7:

Maintenance, Troubleshooting and Product Classification

www.amico.com

17

Advertising