Amico Nema 4 Master Alarm User Manual

Page 17

Advertising
background image

www.amico.com

17

InchHg / KPA / BAR selections

Factory Default - InchHg

For InchHg mode, set the switch #4 to the ON position. The LED indicating InHg located next to the

VACUUM source reading will illuminate.

For KPA mode, set the switch #4 to the OFF position and the switch #9 to the ON position. The LED

indicating KPA located next to the VACUUM source reading will illuminate.

For BAR mode, the KPA indicating source must be changed to BAR by use of a label. Set the switch

#4 to the OFF and the switch #9 to the OFF position. The LED indicating BAR located next to the

VACUUM source reading will illuminate.

COMMON SETTINGS FOR PRESSURE AND VACUUM

Repeat Alarm Enable/Disable

Factory Default - Disable

Disable

Set switch #5 to the OFF position to disable the repeat alarm.

NOTE: When the repeat alarm function is disabled, the alarm will not repeat.

Enable

Enable Mode: (Factory Default 30 min, when enabled).

Set switch #5 to the ON position.

PLEASE NOTE: The Module with the Lowest set Repeat Time is the one that controls the Repeat

Time. For example if one Module is set for 5min and one for 30min and both are Repeat Alarm en-

abled, the Alarm will now Repeat every 5min.

InchHg

KPA

BAR