20

Amico Pipeline

SIGNAL INPUT SELECTION

Factory Default - Normally Closed as per NFPA 99.

The Amico alarm can detect field devices in the Normally Open or Normally Closed position.

For Normally Closed

Set switch #3 to the OFF position.

For Normally Open

Set switch #3 to the ON position.

PLEASE NOTE: The above signal input selection, applies to both alarm systems - the Master/NEMA 4

and the 2 in 1.

MAINTENANCE MODE

Factory Default - Disabled

The Maintenance (or Latch) mode is used to allow hospital personnel to identify loose wiring or faulty source equipment.

By putting the master module into “Latch” mode, any alarms received; even transient ones, will be latched-on so that

maintenance personnel can identify the source of the problem. The Maintenance mode will disable the automatic reset,

if a fault condition has been rectified. The alarm indicator can only be turned-off by pushing the “alarm silence” button

on the annunciator module twice. The “Maintenance” LED will illuminate whenever the maintenance mode is enabled.

1 HR

12

HR

24

HR

Version

3

4

Disable

Set switch #4 to the OFF position.

Enable

Set switch #4 to the ON position.