Amico Nema 4 Master Alarm User Manual

Page 23

Advertising
background image

www.amico.com

23

SYMPTOM

CAUSE

CORRECTIVE ACTION

3. Power light on the annunciator

module is ON but LED’s on other

modules are not on.

4. No audible alarm and LED’s are

not illuminating.

5. Audible signal will not silence.

6. Alarm condition exists but

LED’s are not illuminating.

a. DC power cable is not connected

to the annunciator module.

a. DC power cable is disconnected

or loose.

a. Faulty display module.

b. Connection of the DC power

cable from system power supply

to annunciator module is loose.

c. Faulty annunciator module.

a. Display module not properly

calibrated.

1. Ensure that the DC power cable is

firmly in it’s socket on the annunciator

module.

2. Ensure that the module(s) on the

Frame/Module assembly are all

connected to the ribbon cable.

3. Replace the annunciator module.

1. Ensure that the DC power cable from

the system power supply is connected

to the annunciator module snugly.

2. Depress “TEST” button. If the LED’s

come on and there is no audible,

replace the annunciator module.

If this does not work, try solutions

to problem #2.

1. Disconnect the ribbon cable from

the back of the faulty display module(s)

and replace the module(s).

1. Disconnect the DC power cable from

the annunciator module and then

reconnect. If audible alarm still persists,

replace the System Power Supply unit.

1. Replace annunciator module.

1. Ensure that the system was properly

ordered.

Factory default settings:

Mid Pressure:

Hi Pressure 60 Psi

Low Pressure

40 Psi

Vacuum:

Low Vacuum

12 inHg

High Pressure:

Hi: Nitrogen & Air 195 Psi

Low: Nitrogen & Air

140 Psi