Bio-Rad Gel Doc™ XR+ System User Manual

Page 40

Advertising
background image

40

9. Next you will be prompted to remove your sample from the transilluminator and place the

fluorescent reference plate on the platen for UV-illuminated images. For white light illumination
the white light transilluminator or the White Light conversion screen is used to collect the
reference flat fielding image. In either case, you must remove the sample from the platen.

10. Remove the sample from the transilluminator. Remove any liquid or residue remaining from the

sample platen. Place the fluorescent reference plate on the platen. Make sure that the plate covers
the entire glass surface. Close the door and then click on Continue in the window shown above.

11. The next prompt will direct you to turn ON the UV transilluminator. Press the UV transilluminator

or white light transilluminator button on the hood, as the case may be, and click on Continue.