3 uv transilluminator, 4 lamp replacement – Bio-Rad Gel Doc™ XR+ System User Manual

Page 46

Advertising
background image

46

1. Unplug the main power cable from the power entry module outlet.
2. Use a screwdriver lever on the fuse holder to extract it.
3. Remove the blown fuses and replace them with two new ones (catalog #100-1951).
4. Slide the fuse holder into the power entry module until it snaps in place.

8.3 UV Transilluminator

UV transilluminator without cover

8.4 Lamp Replacement


Attention: The UV filter surface should always be kept clean from the chemical agents used as gel
dyes. Please use protective gloves when touching the UV transilluminator cover.

The lamps must be replaced after every 500 hours of use.

1. Turn off the power.
2. Disconnect the power cord from the Universal Hood II.
3. Remove the four screws located on the left/right sides of the trans cover.
4. Remove the cover with the UV filter by sliding it forward, then lifting it up.
5. Place it down on its backside on a nonabrasive surface so that the glass does not get scratched.
6. Caution: Do not put the UV filter directly on the bench. Wear gloves when touching the lamps.

Remove the lamp by rotating until it becomes loose and the pins come to a vertical position.

7. Remove the lamp. Install the new lamp by rotating so that the pins are horizontal and the lamp is

tight.

8. Reassemble the cover and retighten the screws on both sides.

Lamp replacement Starter replacement