0 changing sheaves – Cooper Instruments & Systems WTM 502 Wire Tension Meter User Manual

Page 11

Advertising
background image

Reso Press

the

Reso softkey and you are prompted to enter a display, or count-by, resolution.

Choose from Low, Medium or High.

Low resolution provides the best stability and makes the display easiest to read. High
resolution provides the finest graduations, but sees greater drift from wire creep and non-
repeatability. If the reading is decreasing over time or differing between measurements
on the same line, lowering the resolution will reduce these effects.

Comm Communication output not supported at this time in the WTM 502

Mode

Press this softkey to set the display mode on power up. Choices are Last*, Temp,
Check, Avg, Peak,
and Force. Use the Sel keys to display your choice and press Enter
to accept it.


Units

Press this softkey to set the following:

• Unit of measure on power up. Choices are Last*, C2, C1, N, kgf, and lbf. Use the Sel keys

to display your choice and press Enter to accept it. C2 and C1 are custom units. If you
choose to have custom units, you are prompted to enter the number of pounds in each
custom unit. The WTM 502 will then automatically calculate correct display for the applied
force.

Enable lbf – Enable of disable the pound-force unit of measure.

Enable kgf – Enable or disable the kilogram-force unit of measure.

Enable N – Enable or disable the N unit of measure.

Enable CUST1 – Enable or disable the Cust1 unit of measure.

Enable CUST2 – Enable or disable the Cust2 unit of measure.

Custom units of measure are handy when working with multi-part lines.

Enable C – Enable or disable Centigrade temperature.

Enable F – Enable or disable Fahrenheit temperature.

Off

Press this softkey to enable or disable the auto-shutdown. If you enable this function you
are prompted to set a period of time in minutes. Next, press the Enter softkey to accept
this value. You are then asked to set the shutdown type; Fixed, No Load, or No Change.
These are described below;

Fixed – The unit will shutdown after the set number of minutes no matter what happens.
No Load – The unit will shutdown after the set number of minutes only if there is no load on
the unit. This prevents shutdown in the middle of line tensioning.
No Change – The unit will shutdown if there has been no keypad activity or change in
tension after the set number of minutes.

ChPwd Press this key and you are prompted to enter a new password to access the

configuration menus. Use the softkeys to scroll in a new password and press the Enter
softkey to accept it.

Default Configuration password is 0. If a new password is lost or forgotten, contact Cooper
Instruments.

Reset

Press this key and you are asked if you wish to reset the system. Press the Yes softkey
only if you want to continue. Press the No softkey to abort this and return to the previous
screen.


4.0 CHANGING SHEAVES

As you use the WTM 502 on different diameter cables you must change to the correct sheave size. To
change sheaves, remove the hex head screws pointed out in Figure 7 below. Replace the sheaves with
the correct letter sheave and reinsert the screws and tighten.

Insure sheaves installed agree with sheaves noted in the Wire calibration.

CF 142

9 V 08/09/05