0 achieving best accuracy, 1 accuracy, 2 calibration to specific wire type – Cooper Instruments & Systems WTM 502 Wire Tension Meter User Manual

Page 12: 1 accuracy 5.2 calibration to specific wire type

Advertising
background image

Exception: Sheaves match the wire diameter of the cable to be measured and alternate calibration is
selected as per section 5.2.

Insure that the wire rope is riding in the groove of all three sheaves.

Do not use the WTM 502 with cable larger than indicated on the sheaves. Overload and damage to the
instrument may result.

Do not mix sheave sizes. This will result in inaccurate measurement and possible overload.

FIGURE 7 CHANGING SHEAVES

5.0 ACHIEVING BEST ACCURACY

5.1 Accuracy

The WTM 502 is an instrument designed to give accuracy that typically exceeds normal requirements for
wire tensioning. You should have an understanding of what factors affect tension measurement accuracy.

5.2 Calibration to Specific Wire Type

While it is best to have the instrument calibrated to the specific wire size(s) and type(s) used, the WTM
502 can often work adequately in other situations. If the best tension accuracy is required, Cooper
recommends that a calibration be performed for that specific wire size and type.

What calibration choice and sheaves should I use?

Situation

Wire calibration

selection

Sheave

selection

Accuracy

Exact wire size and type is

shown in wire list

Description of exact

match

Sheaves noted in list

Best

Wire size is same, but type is

not identical

Description of same wire

diameter

Sheaves noted for that wire size

Good

Wire size is not the same

Closest diameter

Sheaves matching the wire size

being measured

Fair

Contact Cooper Instruments for any additional calibrations you may need.


Do not use the WTM 502 to measure tension for wires if both of the following are true:

CF 142

10 V 08/09/05