Terminal Assignment

Wire

color DIN47100

IEC757

Pin 1

sensor supply +24 V

yellow

red

Pin 2

sensor supply 0 V

green

black

Pin 3

--

--

--

Pin 4

Tx/Rx+, RS485

brown

brown

Pin 5

Tx/Rx-, RS485

white

orange

2

1

3

4

5

connector M16

2

1

3

4

5

mating

connector M16

Connect power supply that is free from interference emission. The supply must not be used as
parallel supply of drives, shields, magnetic valves, etc.
Apply DC according to terminal assignment power supply: 24 VDC ±10 %

Shield
Use shielded cables only. The cable shield must be grounded at the machine
(two examples each depending on the cable, see drawings).



3. Interface

3.1 Interface

data

Type: RS485
Baud rate:

19200

Parity: No
Data bits:

8

Stop bits:

1

Handshake: No
Check sum:

Yes (CRC)

* To avoid Bus collisions, a minimum time lag in responding of 1 ms

is imperative, i.e. the time elapsed between the last bit of the query
and sending the 1st bit of the N 155 response.

Important: Please consider that due to the abovementioned
minimum time lag of 1 ms in responding the Bus switching time

after having sent the last bit must not exceed 1 ms.

Lag time response*:

1...16 ms


3.2 Protocol

The target display operates with ASCII protocol (clear text protocol). Depending on the command, the volume of
the protocol varies between 5 and 17 bytes.

Byte

Hex-Code

Significance

Value range

1

01H

SOH = header token

fix on 01h

2

XXh

Adr = address

00...31dez + 20h Offset (address 00 = 20h)

3

XXh

Cmd = command code defined commands

4-n

[XXh..XXh]

[Data] = data

20h-7Fh

n+1

04h

EOT = footer token

fix on 04h

n+2

XXh

CRC = check sum

00h...FFh