Scilogex DM0142 Clinical-Centrifuge User Manual

Page 12

Advertising
background image

9

z There is a circulating function to increase/decrease the speed values. Rotate the program button

clockwise

to change settings from small → large → maximum → minimum. Rotate the program

button anti-clockwise

to change settings from large → small → minimum → maximum.

2)Set the time

z Press select button

, time value flashes in the time setting mode.

z Rotate the program button

to set running time from 30 seconds to 99 minutes.

z When the time displays HD, this is a continuous running mode.

8.1.2 Start the operation

1)Press button

to start running

z The door must be locked before rotor starts spinning.

z The timer will start once the rotor starts spinning, the screen displays the remaining run time.

2)View and modify the operation programs

z Operation programs can be modified after the centrifuge reaches the set speed.

z Pressing the select button

, returns the display to the program interface and displays setting

programs. Press the select button

to the desired program. When flashing, rotate parameter

button

to modify values. Release the button after 5 seconds, and the centrifuge will return to

normal operation mode and run according to the new value.

z If the set time value has been modified, the operation time is not affected and will continue.

3)Warning display

z If an error occurs during the operation, the centrifuge will brake to stop automatically, and display the

error code on the time/display area. The error code can be checked in the table 10-1, and corrective

actions can be applied accordingly.

8.1.3 End the operation

1)The centrifuge will brake when it reaches the set time or

button is pressed.

z When the rotor stops rotating, the centrifuge will start beeping to alert the operation has finished.

2)Open the door

z The door can be released automatically when the operation has stopped.

z With the door closed, you are able to press the

button to open it.

z After ending the operation, the program will store the setting parameters of this operation, and will

recall these parameters when restarting the program.

3)Open the door and take out the rotor and samples.