Scilogex DM0142 Clinical-Centrifuge User Manual

Page 17

Advertising
background image

14

11.1.2 Available rotors and adapters

Rotor

type

Tubes

Tubes

Per

Rotor

Dimension

(Ф×L

mm)

Adapters

Maximum

speed(rpm)

Radius

cm

Maximum

RCF

(×g)

A12-10P

15ml con

6

17×120

4500

11

2490

1.5-5ml vacu

12

13×82

A10P15

mat

4500 9.8

2218

4-7 ml vacu

12

13×106

A10P15

4500

11 2490

16×75

A10P15

Mat

9.8 2218

8.5-10 ml vacu

12

16×107

A10P15

4500

11

2490

2.7-3

(EU) ml

collection tube

12 11×66

A10P15

mat

4500 9.8

2218

7.5-8.2

(EU) ml

collection tube

12 15×92

A10P15

4500

11 2490

Table 11.1 Rotors and adapters

11.1.3 Notice

z The centrifuge rotor can separate samples with a density lower than 2.0g/ml. If the samples density is

over 2.0g/ml, please calculate allowable speed depending on the following formula.

Allow Speed (rpm)= Maximum speed×(2.0(g/ml)/Sample density (g/ml))

1/2


11.1.4 Autoclaving

A12-10P rotor is made of plastic, cannot be high-pressure sterilization and UV irradiation, only ordinary
sterilization can be used.











Advertising