Caution – Scilogex DM0142 Clinical-Centrifuge User Manual

Page 5

Advertising
background image


2

z If the centrifuge, rotor and accessories that have been contaminated by solutions with toxic,

radioactive or pathogenic materials, clean it according to the decontamination procedure as specified.

z If you require service at site, please sterilize and decontaminate the centrifuge in advance, and then

notify the service center the details of the materials and procedure.

z To avoid electrical shocks, insure hands are dry before handling the power cord or turning on/off the

power switch.

z For safety purposes, do not enter within 30cm around this centrifuge when it is in operation.

z While the rotor is rotating, never release the door lock.

z

Unauthorized repairs, disassembly, or modifying the centrifuge except by our service center are

strictly prohibited.

CAUTION

z This centrifuge must be located on a firm and level table.

z Make sure the centrifuge is horizontal before running.

z Make sure the angle between the door and cover is greater than 70 degrees when opening the door.

z Be careful not put your fingers or hands between the door and cover while the door is open.

z Do not move or relocate the centrifuge when it is running.

z If fluid spills in the rotor chamber, please promptly clean and dry with a dry cloth to avoid sample

contamination.

z Ensure to remove any objects and fragments of the tubes dropped inside the rotor chamber before

running the centrifuge.

z Cautions with rotor

(1) Always check for corrosion and damage on the rotor surface before using it. Do not use the rotor

if an abnormality is found.

(2) Do not set the speed beyond the allowable minimum speed of the rotor kits (rotor and adapters).

Make sure to run it below the allowable maximum speed.

(3) Do not exceed the allowable imbalance.

(4) Use the rotor and tubes within their actual capacities.

(5) If the rotor is attached with a lid, ensure it is tightened before operation.

z If any abnormal condition occurs during operation, please stop it immediately and contact our service

center. Notify the service center is a warning code if displayed.

z Vibrations are likely to damage the centrifuge, contact our service center if abnormality observed.