Instructions for the rotor and tubes, 1 rotor instructions – Scilogex DM0142 Clinical-Centrifuge User Manual

Page 16

Advertising
background image

13

10.2.2 In the case of power outage

The door cannot be opened automatically if there is a power outage. It is available to be opened manually

as follows.

1)Ensure if the rotor has stopped rotating.

z Listen carefully to ensure no rotating sound can be heard.

2)Insert a screw driver into a hole to open door.

z Holes are located on the left and right sides of the unit.

z Insert a screw driver into the two holes and push forward to release the door.

11. Instructions for the rotor and tubes

CAUTION

z Read the instructions thoroughly, to properly load and use rotor.

z Do not exceed the allowable maximum speed of rotor、tube and adapters etc. Ensure the allowable

maximum speed of adapters is lower than the rotor’s maximum speed.

11.1 Rotor instructions

11.1.1 Rotor structure

Figure 11-1 The rotor structure

adapter mat adapter screw nut rubber cap

Rotor body mat Spindle hole