Scilogex DM0142 Clinical-Centrifuge User Manual

Page 14

Advertising
background image

11

1)Centrifuge

z If the centrifuge is exposed to ultraviolet rays for a long time, the color of the door may be changed or

the label may be peel off. After using, cover the centrifuge with a piece of cloth to protect it from

direct exposure.

z If the centrifuge needs cleaning, clean it with a cloth or sponge moistened with a neutral detergent

solution.

z Sterilize the centrifuge by wiping with a cloth moistened with 70% ethanol solution.

2)Rotor chamber

z If the rotor needs cleaning, clean with cloth or sponge moistened with a neutral detergent solution.

Sterilize the centrifuge by wiping with a cloth moistened with 70% ethanol solution.

3)Drive shaft

z We recommend regular maintenance for drive shaft. You can wipe the drive shaft with soft cloth, and

then apply a thin coat of silicon grease.

4)Door

z Clean and sterilize the door using the same method as the section(1)above.

5)Rotor

z To prevent corrosion, remove the rotor from rotor chamber. If not in use for a long term, then detach

the rotor lid and turn upside down to dry the tube holes and keep clean.

z For sample leaks in the rotor, rinse the rotor with water. Apply a thin coat of silicon grease to the rotor

when it is completely dry.

z The rotor should be checked every 3 months to ensure the tube and rotor holes keep are clean and

apply a thin coat of silicon grease.











CAUTION

z Do not directly pour water, neutral detergent or disinfectant solution into the rotor chamber, otherwise

fluids may leak into the drive units and cause corrosion or deterioration to the bearings.