10. Troubleshooting

10.1 Possible problems and solutions

This centrifuge has a self-diagnostic function. If a problem occurs, an error/warning code will be displayed

on the time display screen, and the operator can determine the malfunction with the alarm code below.
Symptom Causes

Solutions

Nothing appears on the screen when

the POWER is turned on.

·

Building power circuit breaker

trips.

·

Remove the trouble and turn

on the POWER.

Abnormal vibration

·

Rotor do not match with

spindle

·

Samples are imbalance

·

Install again the rotor

·

Weighting scales, install

symmetrically

Alarm code a

ppe
ared

on t
he tim

e

disp

la

y scree

n

E-02

Door fault

·

The door opened in running.

·

The

button is pressed

while the door opening.

·

Close the door immediately.

·

Close the door,and then start

to operate.

E-06

Set wrong speed

·

The setting speed exceed the

allowable range.

·

Modify the speed value.

E-10~86

·

Read the service manual

·

Contact with service center

Table 10-1 Possible problems and solutions

z

Alarm codes E-1~E-9 are related to incorrect operation/programming. You can continue running the

centrifuge after implementing corrective procedures.

10.2 How to open the door

10.2.1 In the case of

power on

1) Turn on the power switch, release the door automatically.

2) The door will be released automatically once the operation is finished.

3) It is available to release the door by press

button once the rotor stops.

CAUTION

z The door just can be opened while the power is on and rotor stops rotating.