30

590 AG Indicator Operators Manual

5.1.2

Reporting Commands

Reporting commands (see Table 5-2) send specific
information to the EDP port. These commands can be
used in both setup mode and normal mode.

Command

Function

DUMPALL

List all parameter values

VERSION

Write 590 AG Indicator software version

RS

Reset the indicator

P

Write current displayed weight with units
identifier.
The P command returns the current
displayed weight value to the EDP port. If
the indicator is in an underrange or overload
condition, the weight value is replaced with
###### (overload) or %%%%%%
(underrange).

S

Write one frame of stream format

Table 5-2. EDP Reporting Commands

5.1.3

The RESETCONFIGURATION Command

The RESETCONFIGURATION command can be used
to restore all configuration parameters to their default
values. Before issuing this command, the indicator must
be placed in setup mode or test mode (press and hold
setup switch for approximately four seconds to enter
test mode).

Sending this command, followed by the RS (RESET)
EDP command, is equivalent to using the DEFLT
function on the TEST menu. See Section 8.7 for more
information about test mode and using the TEST menu.

NOTE:

All load cell calibration settings are lost when

the RESETCONFIGURATION command is run.

5.1.4

Parameter Setting Commands

Parameter setting commands allow you to display or change the current value for a particular configuration
parameter (Table 5-3 through Table 5-8).

Current configuration parameter settings can be displayed in either setup mode or normal mode using the following
syntax:

command<ENTER>

Most parameter values can be changed in setup mode only. Use the following command syntax when changing
parameter values:

command=value<ENTER>

where

value

is either a number or a parameter value. Use no spaces before or after the equal (=) sign. If you type an

incorrect command, the display reads

??

. Changes to the parameters do not take effect until you exit setup mode.

For example, to set the motion band parameter to 5, type the following:

MOTBAND=5D<ENTER>

Table 5-3. CONFIG EDP Commands

Command

Description

Values

GRADS

Graduations

1–999 999

ZTRKBND

Zero track band

OFF, 0.5D, 1D, 3D

ZRANGE

Zero range

1.9%, 100%

MOTBAND

Motion band

1D, 2D, 3D, 5D, 10D, 20D, OFF

OVRLOAD

Overload

FS+2%, FS+1D, FS+9D, FS

SMPRAT

Sample rate

15HZ, 30HZ, 3.75HZ, 7.5HZ

DIGFLTR1
DIGFLTR2
DIGFLTR3

Digital filtering

1, 2, 4, 8, 16, 32, 64

DFSENS

Digital filter cutout sensitivity

2OUT, 4OUT, 8OUT, 16OUT, 32OUT, 64OUT, 128OUT

DFTHRH

Digital filter cutout threshold

NONE, 2DD, 5DD, 10DD, 20DD, 50DD, 100DD, 200DD, 250DD