590 AG Indicator Operators Manual

1.0

Introduction

The 590 AG Indicator is a single-channel digital weight indicator housed in a NEMA 4X/IP66-rated stainless steel
enclosure. The indicator front panel consists of a one inch (25 mm), six-digit, liquid crystal (LCD) display and 19
button keypad. Features include:

Drives up to four 350

or eight 700 load cells

Supports four and six wire load cell connections

Electronic data processing (EDP) port for full duplex, RS-232 communications at up to 19200 bps

Powered by six C batteries for complete portability

AC adapter for 115 or 230 VAC power

Configurable standby mode limits indicator power consumption when scale is inactive

The 590 AG Indicator is NTEP-certified for Classes III and III L at 10,000 divisions. See Section 8.8 for detailed
specifications.

1.1

Operating Modes

The 590 AG Indicator has several modes of operation:

Normal (Weighing) Mode

Normal mode is the “production” mode of the indicator. The indicator displays gross weight, using the LCD
annunciators described in Section 1.3 to indicate scale status and the type of weight value displayed. Once
configuration is complete and a legal seal is affixed to the back of the indicator, the 590 AG Indicator can operate
only in normal mode.

Setup Mode

Most of the procedures described in this manual require the indicator to be in setup mode, including configuration
and calibration.

To enter setup mode, remove the large fillister head screw from the enclosure backplate. Insert a screwdriver or a
similar tool into the access hole and press the setup switch once. The indicator display changes to show the word

CONFIG

.

Test mode

Test mode provides a number of diagnostic functions for the indicator. Like setup mode, test mode is entered using
the setup switch. See Section 8.7 for more information about entering and using test mode.

Standby mode

Standby mode is a configurable low-power mode used to extend battery life when the indicator is inactive. In
standby mode, the LCD display shows the word

STNDBY

; press any key to exit standby mode. See Section 3.2.5 for

information about configuring standby mode.

Advertising