Calibration Procedure

See your indicator operation manual to determine indicator calibration procedures.
We recommended that you “exercise” the scale before calibration to be certain that everything is seated. Load the
scale to near capacity two or three times. Then, with no load on the scale, place the indicator in its calibration mode
and perform a zero calibration. Place test weights equal to 70% – 80% of the scale capacity on the platform. If you
use several weights, distribute them evenly around the platform. Perform a span calibration.
Remove the test weights and check the zero reading.
Repeat the calibration process if necessary.