Figure 4-3. foot pad - side view, Dimension shown in figure 4-3 – Rice Lake AutoLift HE Low-Profile Lift User Manual

Page 16

Advertising
background image

12

RoughDeck HE AutoLift

Figure 4-3. Foot Pad - Side View.

Route the load ell cables near each corner so the cable is free from possible contact with each foot. Hold the cable
in position with the adhesive-backed cable ties supplied in the scale kit.
Do not cut excess load cell cables. Make a strain relief loop before routing the cables through the conduit. Coil
extra cable before it enters the junction box. After coiling excess cable, pass the end of the load cell cable through
the cable fittings in the NEMA 4X junction box.
Corner correction, trimming, and calibration are necessary after replacing a load cell. Follow the instructions in
Section 3.2 and Section 3.3 on page 10.

2K – 10K models: 1.625"
20K models: 2.16"
30K models: 1.66"