background image

14

RoughDeck HE AutoLift

4.4

Troubleshooting Guide

Symptom

Probable Cause

Solution

System does not
operate--no display

Power disconnected

Check and reconnect

Indicator fuse blown

Replace fuse, check for cause

Interface cable cut or disconnected

Repair

Signal leads incorrectly installed at indicator

Install according to indicator installation manual

Display stays at zero

Faulty load cell connections

Check cable connections in J-box at indicator

Faulty indicator

Service the indicator

Erratic weights

Vibration near scale

Remove source of vibration or move scale

Platform not level to within 1/4”

Level the platform by adjusting feet or shimming

Load cell or cable water damage

Replace load cell or cable

Debris under load cells or platform

Clean

Indicator faulty

Use simulator to test indicator stability; service
indicator

Consistently high or low
weights

Indicator not properly adjusted to zero

Zero the indicator according to operator’s manual

Platform binding

Adjust clearance for free platform movement

Indicator not calibrated

Calibrate according to indicator’s manual and
Section 3.3 on page 10

Faulty load cells

Test and, if necessary, replace load cells

Feet touching deck’s underside

Adjust feet downward to provide clearance

Table 4-2: Troubleshooting Guide.