1 programming mode menu – Rice Lake Bariatric Wheelchair Scale (SB-1150) User Manual

Page 17

Advertising
background image

SB-1150 Installation & Operation Manual - Scale Configuration

13

7.1

Programming Mode Menu

Various parameters can be set while in programming mode.
Those parameters are shown in Figure 7-3.

Figure 7-3. Programming Mode Menu Structure

Table 7-1 lists the various display messages and sequence when setting up the scale.

Parameter

Description

Choices

Steps

FULL

Full capacity of the scale

Value (1000 lb)

The display toggles between a numeric value and FULL.
If you don’t want to change the value, press the BMI key to
move to the next setting. Example: from FULL to LOAD. If you
want to change the value, use the following steps.

1.

Press the ENTER key to change value starting with
the right most digit.

2.

Use the Up/Down arrow keys to increment/
decrement numbers.

3.

Press the BMI key to move to the left.

4.

Use the Up/Down arrow keys to increment/
decrement numbers.

5.

Press the BMI key again to move to the left

6.

Use the Up/Down arrow keys to increment/
decrement numbers.

7.

When done press the ENTER key to move to the
next parameter (LOAD).

Table 7-1. Configuration Mode Menu

Power Off

Power On-Off

START

Kg-Lb Key

0.000

Software Version

Round xx

Full xxx

A TOLXX

Drange

At

Press and
hold

A LEN X

MESS

A TImE XX

BAUD XX

OP1=X

SAVE

T-OFFX

DONE

OP2=X

OP4=X

OP3=X

OP5=X

OP7=X

OP6=X

OP8=X

Press and
hold the
Kg-Lb key
to advance
to option
2 through
option 8.

= DEF =

Software ID

SET=X

OP9=X

OP10=X