3 load cell replacement – Rice Lake Bariatric Wheelchair Scale (SB-1150) User Manual

Page 31

Advertising
background image

SB-1150 Installation & Operation Manual - Maintenance

27

11.3

Load Cell Replacement

Replacement load cells can be ordered from Rice Lake Weighing Systems. To replace a load cell, lift the scale

with a chain and remove foot, then remove the defective load cell. Disconnect load cell cable from the junction

box and cut cable ties. When the cable is freed, pull cable out of the scale frame channels.

Figure 11-1. Load Cell Assembly

Follow the directions given below to install new load cells.
Lay out the four load cells near the corners where they are to be installed. Thread the cable from each load cell

through the conduit tubing in the frame and into the junction box according to the wiring diagram in Figure 9-3.
Note that in Figure 11-2 both the scale and the junction box are viewed from the bottom. To verify correct load

cell/junction box terminal matching, see the numbers on the terminals inside the junction box and the corner

numbering diagram in Figure 11-4 28.

Figure 11-2. Bottom View of Scale

Check that the threaded holes for the load cell screws are free of debris. Use compressed air to blow out holes if

necessary. Position load cells with alignment arrows pointed up toward the deck and loosely install the hex head

cap screws provided, as shown in Figure 11-1. If the base is used with a pit frame or access ramp, position the

load cell to maintain the dimension shown in Figure 11-3. With the torque wrench, tighten all bolts to 75 ft-lb.

CELL 4

CELL 3

CELL 2

CELL 1

1

4

2

3

Advertising