0 summit sb-1150 scale specifications – Rice Lake Bariatric Wheelchair Scale (SB-1150) User Manual

Page 33

Advertising
background image

SB-1150 Installation & Operation Manual - Summit SB-1150 Scale Specifications

29

12.0

Summit SB-1150 Scale Specifications

Electrical Grounding

For systems where the scale is connected to a 115 VAC circuit, the indicator must be directly connected to an earth ground with
a ground interface cable of no more than 3W resistance throughout its length.

Load Cell Excitation

Rated Excitation: 10 VDC
Maximum Excitation: 15 VDC

Grade Level Requirements

The supporting surface for the four feet of the scale must be level within 1/4 inch of horizontal.

Safe Static Overloading Capacity

Maximum: 150% of scale capacity

Nominal Scale Height

3.0 inch (76 mm)

Indicator Specifications

Power

120 VAC-9VDC-50Hz / 230 VAC-9VDC-50Hz

Battery Type

Sealed lead acid battery

Battery Use

75 hours
Automatic power-off can be configured

Environmental

Operating Temperature

50 to +104°F (14 to 40°C)

Storage Temperature

32 to 158°F (0 to 70°C)

Humidity
85% relative humidity

Certifications and Approvals

RoHS Compliant