4 assembly, 5 electrical interface to indicator – Rice Lake Bariatric Wheelchair Scale (SB-1150) User Manual

Page 8

Advertising
background image

4

SB-1150 Installation & Operation Manual

3.4

Assembly

For load cell protection during shipping, the scale feet are shipped separately from the floor scale.
Screw one foot into each load cell and turn all the way in until the foot touches either the load cell or the

underside of the deck. Then unscrew each foot three complete turns.
Place a spirit level on the deck. Adjust any “high” corners not in contact with the floor by further unscrewing the

feet on those corners until they just contact the floor surface. When all feet are in contact with the floor, check the

deck with the spirit level to be sure the scale is within 1/4 inch of level.

3.5

Electrical Interface to Indicator

Ten feet of 4-wire cable used to connect the scale to the weight indicator is supplied with each scale. The junction

box is easily accessible through an access plate located on the side of the Summit SB-1150. Use the following

steps to wire up the junction box.

1. Remove the two #10 x 3/8" screws.
2. Slide the junction box assembly out of the deck.
3. Open and remove the top of junction box.
4. Push the cable end into the junction box through a cord grip.
5. Connect the wires to the indicator terminal (Figure 3-2) as shown in Table 3-1.
6. Pull out excess and tighten the cord grip to hold the cable snugly.

Figure 3-2. Junction Box Wiring Diagram

Cable Color Code

Junction Box

Red

+ Excitation

Black

- Excitation

Green

+ Signal

White

- Signal

Bare

Shield

Table 3-1. Junction Box Connections