Rice Lake Chair Scales User Manual

Page 14

Advertising
background image

10

Chair Scale Technical and Operating Instructions

The following table lists the various display messages and sequence when setting up the scale.

NNOTE:

<-> means the display toggles between the two values.

Step

Function

Display

Available Parameters

1

With the scale turned
off, remove the
confiiguration jumper
as shown in
Figure 4-1.

Allows the scale to enter into either
configuration or calibration mode.

2

With the scale turned
off, simultaneously
press the On-Off/
Zero and Kg-Lb key.

Enters into
programming mode

StArt

Scale automatically advances to Step
3.

3

Indentifies the
software ID

IdEnt<->11305

Press the REWEIGH key to advance
to the next step.

4

Identifies software
version

Id<->11387

Press the REWEIGH key to advance
to the next step.

5

Allows selection of
decimal point setting

dOt<->000.0
Default = 0.0

To change the position of the decimal
point, press the Kg-Lb key to toggle
through the various options. To
advance to the next step, press the
REWEIGH key.

6

This indicates the
maximum allowed
weight. Any weight
above this value will
cause StOP to appear
on the display while in
the weighing mode.

FULL<->XX.XXX
Default= 600 Lb

Note: This value is
dependent on the model you
have. Refer to the serial
number label on your scale
and set appropriately.

Use the numeric data entry (See
“Numeric Data Entry” on page 9.
) to
change the value. To advance to the
next step, press the REWEIGH key.

7

This limit defines the
start of the weighing
threshold to
compensate for a key
press during the Tare
function.

Limit<->0.0

Use the numeric data entry (See
“Numeric Data Entry” on page 9.
) to
see the optimum weight value of 5 lbs.
To advance to the next step, press the
REWEIGH key.

8

Display divisions

rOUnd<->XXXX
Default = 0.2 Lb

To change the display divisions, press
the Kg-Lb key to toggle through the
various options.
0.2, 0.5, 1.0, 2.0, 5.0, 10.0, 20.0.
To advance to the next step, press the
REWEIGH key.

9

Double ranges limit

drAnGe <-> 0

This parameter is preset from the
factory. To advance to the next step,
press the REWEIGH key.

10

Weight algorithm initial
tolerance

A tOL <-> 10

This parameter is preset from the
factory. To advance to the next step,
press the REWEIGH key.

11

Weight algorithm initial
exponent

A LEn <-> 8

This parameter is preset from the
factory. To advance to the next step,
press the REWEIGH key.

12

Weight algorithm
maximal exponent

A t INE <-> 10

This parameter is preset from the
factory. To advance to the next step,
press the REWEIGH key.

Table 4-1. Configuration Mode Menu

Advertising