Chair Scale Technical and Operating Instructions - Troubleshooting and Testing

21

7.0

Troubleshooting and Testing

Refer to the following instructions to check and correct any failure before contacting service personnel.

Symptom

Possible Cause

Corrective Action

Scale does not turn on

Dead battery

Connect the scale to a power source.

Faulty electrical outlet

Use a different electrical outlet.

Bad power supply

Replace adaptor.

Questionable weight or the scale
does not zero

External object is interfering with the
scale

Remove the interfering object from the scale.

Display did not show 0.0 before
weighing

Help the patient off the scale, zero the scale and
begin the weighing process again.

Scale is not placed on a level floor

Ensure the scale is level and begin the weighing
process again.

Scale is out of calibration

Check the weight with a known weight value.

Improper tare

Place the tare item on the scale. Press
REWEIGH. Once the weight of the item is
displayed, press TARE. Place the patient back on
the scale. Press the REWEIGH button again.

Weighing is performed but the display
shows WEIGH and REWEIGH every
few seconds; the weighing process
takes too long and no weight is
displayed

The patient is not sitting still

Ask the patient to be still.

The display shows a STOP message

The load on the scale exceeds the
capacity of the scale

Remove the excess weight and use the scale
according to manufacture’s specs.

The display shows LO Bat message

The battery is low

Recharge the battery.

The display shows Err message as
detailed in the table below

Err 2

Low saturation state (low A/D)

The load cell is not connected properly. Check
the cables and mechanical connections. If the
problem persists, replace the set of load cells.

Err 3

High saturation state (high A/D)

See Err 2

Err 6

Unstable weight. Cannot calibrate

Check the load cells’ mechanical surroundings
and see that nothing touches them and that the
cables are properly welded.

Err 7

Mathematical error; division by zero.
Cannot calculate calibration factor.

This error will show when trying to calibrate the
unit with no calibration weight on the unit.

Calibration or configuration settings
are messed up

Pushed the wrong keys

Reset to factory defaults. See Page 9 for
instructions.

The display shows a Con En message The configuration jumper is not in

place.

Install the configuration jumper on both pins. (see
page 4 for more information).

Table 7-1. Troubleshooting Table for the Rice Lake Scale Line