Rice Lake Chair Scales User Manual

Page 15

Advertising
background image

Chair Scale Technical and Operating Instructions - Scale Configuration

11

13

Message style on
weight algorithm

MESS <-> WEIGH

This displays the message that will
show on the indicator display. To
change the message, press the Kg-Lb
key.
WEIGH,
LIVE,
- - - - - -
To advance to the next step, press the
REWEIGH key.

14

This allows for setting
the baud rate of the
RS-232 connection

BAUd <->
Default - 9600

To change the baud rate, use the
numeric data entry (See “Numeric Data
Entry” on page 9.)
To advance to the next step, press the
REWEIGH key.

15

Optional features
Option 1 allows the
selection of unit of
measure (UOM) in
calibration and
programming.

OP1 = 1

Hold the Kg-Lb key for browsing
through the options. Double click on
the Kg-Lb key to change the options
value or
to advance to the next step, press the
REWEIGH key.
OP1 = Unit of measure (UOM) in
calibration and programming.
0=Kg
1=Lb

Option 2 allows the
scale to work only in
Kg.

OP2 = 0

Hold the Kg-Lb key for browsing
through the options. Double click on
the Kg-Lb key to change the options
value or
to advance to the next step, press the
REWEIGH key.
0=Disable
1= Enable
This option works only if Option 3 is
disabled

Option 3 allows the
scale to work only in
Lb.

OP3 = 0

Hold the Kg-Lb key for browsing
through the options. Double click on
the Kg-Lb key to change the options
value or
to advance to the next step, press the
REWEIGH key.
0=Disable
1=Enable

Option 4, the scale
must be stable to
show a Kg or Lb
weight reading. You
can enable or disable
this.

OP4 = 0

Hold the Kg-Lb key for browsing
through the options. Double click on
the Kg-Lb key to change the options
value or
to advance to the next step, press the
REWEIGH key.
0=Disable
1=Enable
Note: won’t show lb or kg until the
scale is at a standstill.

Step

Function

Display

Available Parameters

Table 4-1. Configuration Mode Menu