3.5.2 Parts Counting (Cont.)

Pressing the

Function button displays the APW on the secondary display. The example shows an

APW of 2.000g. Press the

Function button again to return to Parts Counting Mode.

To establish a new APW from Count Mode, press

Mode until COUNT appears. Press No to incre-

ment,

Back to decrement, Exit to leave without saving and press Yes to accept.

APW Optimization
The only adjustment for Parts Counting is APW Optimization. APW Optimization is set On or Off in the Mode menu when Parts Counting is
turned On. When APW Optimization is set on, the APW is automatically optimized. This results in more accurate parts counting. The factory
default is On.
When a new APW has been established, APW Optimization occurs when the number of pieces added to the pan are between one and three
times the number already on the pan. The secondary display will momentarily indicate APW OPT.

3.5.3 Percent Weighing

Use the Percent Weighing Mode to measure the weight of a sample as a percentage of a pre-estab-
lished reference weight. Refer to Section 3.7.5 to enable Percent Weighing. Press and hold

Mode

button until MODE PERCENT is displayed

To Establish a Reference Weight
PERCENT is displayed with CLEAR REF? (blinking). Press the

No button to use the stored reference

weight and begin Percent Weighing.

Press

Yes to continue. PUT SAMPLE (blinking) will be shown on the display.

Place the sample on the pan and press the

Function button. The balance indicates the sample

weight on the primary display as 100%. The actual sample weight is shown on the secondary
display. This example illustrates 20 grams as the reference weight.

Remove the reference weight and place the item to be compared on the pan. The balance indi-
cates the actual percentage on the primary display. The secondary display indicates the weight in
the selected measuring unit.

Pressing the

Function button momentarily displays the reference weight on the secondary display.

Pressing the

Function button a second time will return to the actual weight.

Establish New Reference Weight
Press and hold the

Mode button until PERCENT is shown on the secondary display. PUT SAMPLE

will blink on the lower display. Add new reference weight and press

Function button to accept new

100% reference.