Rice Lake Discovery Series Semi-Micro and Analytical, Ohaus User Manual

Page 38

Advertising
background image

EN-34

Ohaus Discovery

3.7.2 Calibration (Cont.)
Calibration Adjust

Calibration Adjust may be used to adjust the result of the internal calibration by ±100 divisions.

Note: Before making a calibration adjustment, perform an internal calibration. To verify whether an adjustment is needed, place a test mass
on the pan and note the difference (in divisions) between the nominal mass value and the actual reading. Refer to Calibration Masses Table
and use the highest value from the Span Calibration Points column as the test mass. If the difference is within ±1 division, calibration adjust-
ment is not required. If the difference exceeds ±1 division, calibration adjustment is recommended. Following a calibration adjustment, repeat
the internal calibration and verification procedures.

To perform a calibration adjustment, press the

Cal button and release it when MENU CALIBRATE is

displayed.

Press

Yes to enter the Calibration sub-menu. Press No until CAL ADJ is displayed.

Press

Yes to enter the CAL ADJ menu item and view the current setting. If the actual reading was

less than the nominal mass value, a positive adjustment is required.

Press

No until the setting matches the difference noted earlier in the procedure. If the actual read-

ing was greater than the nominal mass value, a negative adjustment is required.

Press

Back until the setting matches the difference noted earlier in the procedure. Press Yes to

accept and store the setting.

CALIBRATION MASSES TABLE

Model

Span Calibration Points (1)

Weight Class

DV114C

100g

ASTMI Class I

OIML E1

DV214C

200g

ASTMI Class I

OIML E1

DV314C

300g

ASTMI Class I

OIML E1

DV215CD

200g

ASTMI Class I

OIML E1

80104130.indd 34

11/7/2006 9:22:44 AM