Ohaus Discovery

Automatic Internal Calibration
Before Automatic Internal calaibration can be used, it must be turned on first in Menu Calibrate.
With the balance on, press and hold the

Cal button until MENU CALIBRATE is displayed, then

release the

Cal button, CALIBRATE is (blinking).

Press the

Yes button to enter calibration, CAL AUTO CAL is displayed.

Press the

Yes button to select Autocal. After pressing the Yes button, AUTOCAL SET OFF is dis-

played.

Pressing the

No button will change the display to AUTOCAL SET ON, then press the Yes button.

With Autocal set on, exit the Calibrate Menu. The balance may now be calibrated automatically
from the weighing mode.

Internal calibration can be accomplished without using an external weight. With the balance on,
press the

Cal button, balance begins to calibrate.


Press

Exit to abort calibration. When internal calibration is finished, the balance returns to the cur-

rently selected mode.

Automatic internal calibration is a feature included with all Discovery models. AutoCal will au-
tomatically calibrate the balance each time a change in temperature significant enough to affect
accuracy, or after every 11 hours.

Span Calibration
With the balance on, press and hold the

Cal button, CALIBRATE (blinking) is displayed.

Press the

Yes button to enter calibration.

Press the

No button at AutoCal.

Press the

Yes button to initiate span calibration.

First the zero reading is taken. Next the display shows the default span calibration weight value.

Press

No to change to an alternate calibration weight. Press Yes when the desired calibration

weight is displayed.

Place the specified calibration weight on the pan. After calibration, remove the weight from the
pan.

To abort calibration at any time, press the

Exit button.