Rice Lake Discovery Series Semi-Micro and Analytical, Ohaus User Manual

Page 26

Advertising
background image

EN-22

Ohaus Discovery

3.5.9 Density (Cont.

Before density measurements can be made, the density mode of operation must be set up in the Menu, Mode Sub-menu. It is in the Mode
Sub-menu where solid, porous, water or auxiliary liquids are selected.

After the basic parameters have been set, the balance density operation is further determined in the APPL DENSITY menu. This menu allows
the setting of Density, Temperature, Dry Weight of Porous Material, Sinker Volume and Density of Oil.

To Perform a Density Measurement
Refer to paragraph 3.7.5 for density set-up procedures.

A. Measuring the Density of a Solid Using Water
The following menu items should be turned on: SOLID, WATER.
Press and hold

Mode until MODE DENSITY appears, DENSITY CHG PARAM? is displayed.

Press the

Yes button, TEMP 20.00 C (blinking) is displayed. Measure the temperature of the

water with a precision thermometer and enter the temperature using the

Yes, No and Back buttons.

When the last digit is entered, press the

YES button to save.

The balance indicates 0.0000 IN AIR (blinking). Weigh the sample in air and press the

Function

button.

Now, weigh the sample in water (liquid) and press the

Function button. The balance displays the

result in grams per centimeter (g/cc). Typical samples are shown.

B. Measuring the Density of a Buoyant Material Using Water
The following menu items should be turned on: SOLID, WATER.
Follow the same procedure in A except refer to Buoyancy Sample weighing illustrations and place
sample as shown.

To Establish New Density Parameters
Press and hold the

Mode button until DENSITY is shown on the secondary display.