Rice Lake Discovery Series Semi-Micro and Analytical, Ohaus User Manual

Page 21

Advertising
background image

EN-17

Ohaus Discovery

3.5.7 Totalize

Totalize measures the cumulative weight of items. Refer to Section 3.7.5 to enable totalizing.

Press the

Mode button, then release when MODE TOTALIZE is displayed. The primary display

shows the current weight.

To Establish Totalize Parameters
The primary display shows TOTAL and the secondary display shows CHG PARAM? (blinking). At
this point you can select either a manual or automatic mode. Press the

Yes button. MODE is

shown in the primary display with SET MANUAL in the secondary display (blinking). Press the

No

button to change mode and

Yes to accept.

Manual Mode
When entering the manual mode, the primary display shows the current weight, the secondary dis-
play shows the current totalized weight. Pan must be clear and stable before beginning process.
secondary display will blink when balance is ready to begin totalizing weights.

Place the item on the pan. Press the

Function button to add the weight of the item to the total. The

new total is shown on the secondary display when stability is reached.

Remove the item from the pan. Balance must be stabilized at zero before next item can be placed.
The primary display indicates the current weight on the pan. The secondary display indicates the
total accumulated weight.

Add the next item and press the

Function button. The display shown indicates a cumulative weight

of 70 grams. Remove item from pan and repeat process for each new item.

Automatic Mode
After entering the automatic mode, place the item on the pan. The weight of the item is shown on
the primary display. The secondary display will show BUSY (blinking) after the item is added until
the balance is stabilized and then shows the new total on the secondary display. Remove the item
from the pan. Balance must be stabilized at zero before next item can be placed, then add the
next item.

To Establish New Totalize Parameters
Press and hold the

Mode button until TOTALIZE is shown on the secondary display.

3.5.8 High Point

The High Point Mode allows the user to capture and store the highest (stable) weight in a series of
weight measurements. Refer to Section 3.7.5 to enable High Point mode. Press the

Mode button

until HI POINT is shown on the secondary display. When set On, the primary display will indicate
the current weight reading, the secondary display will indicate the highest weight reading. The
example indicates a current weight of 50.0000 g and a high point reading of 55.0000 g. Press
Function button to view lowest reading in memory.

To reset High Point and clear memory, press and hold

Mode until High Point is shown on lower

display.

80104130.indd 17

11/7/2006 9:22:34 AM

Advertising