3 capacity x readability, 4 communication – Rice Lake Discovery Series Semi-Micro and Analytical, Ohaus User Manual

Page 49

Advertising
background image

EN-45

Ohaus Discovery

5.3 Capacity x Readability

Unit

DV114C

DV214C

DV314C

DV215CD

milligram

110000.0000 x 0.1

210000.0000 x 0.1

310000.0000 x 0.1

81000.000 x 0.01

gram

110 x 0.0001

210 x 0.0001

310 x 0.0001

81.000000 x 0.00001

ounce

3.8801356 x 0.000005

7.4075316 x 0.000005

10.9349276 x 0.000005 2.857191 x 0.0000005

carat

550.000000 x 0.0005

1050 x 0.0005

1550 x 0.0005

405.000000 x 0.00005

pennyweight

70.731639 x 0.0001

135.033129 x 0.0001

199.334619 x 0.0001

52.084207 x 0.00001

ounce troy

3.5365825 x 0.000005

6.7516575 x 0.000005

9.9667325 x 0.000005

2.604211 x 0.0000005

grain

1697.5596 x 0.002

3240.7956 x 0.002

4784.0316 x 0.002

1250.02116 x 0.0002

tael (Hong Kong)

2.9388975 x 0.000005

5.6106225 x 0.000005

8.2823475 x 0.000005

2.1640973 x 0.0000005

tael (Singapore)

2.9101017 x 0.000005

5.5556487 x 0.000005

8.2011957 x 0.000005

2.1428931 x 0.0000005

tael (Taiwan)

2.9333337 x 0.000005

5.6000007 x 0.000005

8.2666677 x 0.000005

2.160000 x 0.0000005

momme

29.333337 x 0.00005

56.000007 x 0.00005

82.666677 x 0.00005

21.600003 x 0.000005

5.4 Communication

The balance is equipped with an RS232 interface. Connecting the balance to a computer enables you to operate the balance from the com-
puter, as well as receive data such as displayed weight.

5.4.1 Commands

Commands listed in the following table will be acknowledged by the balance. The balance will return “ES” for invalid commands.

Command Function

IP Immediate Print of displayed weight (stable or unstable).

P Print displayed weight (stable or unstable).

CP Continuous Print. 0P ends Continous Print.

SP Print displayed stable weight.

SLP Auto Print stable non-zero displayed weight.

SLZP Auto Print stable non-zero weight and stable zero reading.

xP Interval Print x = Print Interval (1-3600 seconds) 0P ends Interval Print.

H Enter Print Header Lines

Z Same as pressing Zero Key

T Same as pressing Tare Key.

xT Establish a preset Tare value in grams. X= preset tare value in grams. To clear tare, enter 0 for x.

PT Prints Tare weight stored in memory.

PM Print current mode (weighing mode).

M Scroll to the next enabled mode.

PU Print current weighing unit.

U Scroll to the next enabled unit.

OFF Turns balance OFF.

ON Turns balance on.

PSN Print Serial Number.

PV Print Version: name, software revision and LFT ON (if LFT is set ON).

x# Set PC reference weight (x) in grams. (Must have an APW stored.)

P# Print PC reference weight.