Rice Lake Intercomp CS 5000 - Users Manual User Manual
Page 15
CS5000, Users
Rev V, January, 2008
15
2: The IZSM (initial zero setting mechanism on power up) must be within +/- 10% of the
zero obtained at calibration.
3: The push-button zero and AZSM can only operate within +/- 2% of the IZSM.
4: The overcapacity point is 103% of capacity above the IZSM.
9. The scale shows “EE-05
EE-05
EE-05
EE-05”. Press the ZERO and PRINT keys. Enter the desired
graduation. Use the following table to select a graduation value. (Press the ZERO
and PRINT keys to enter this value.)
Setting
Count by
in lb
Count by
in kg
0
100
50
1
50
20
2
20
10
3
10
5
4
5
2
5
2
1
6
1
0.5
7
0.5
0.2
8
0.2
0.1
9
0.1
0.05
10
0.05
0.02
11
0.02
0.01
NOTE: To meet stated accuracy statements, it is suggested that you choose your
graduation setting from the table below:
If your capacity is:
Set your graduation to:
500 lb / 250 kg
0.2 lb / 0.1 kg (8)
1000 lb / 250 kg
0.5 lb / 0.2 kg (7)
2,000 lb / 1,000 kg
1 lb / 0.5 kg (6)
5,000 lb / 2,500 kg
1 lb / 0.5 kg (6)
10,000 lb / 5,000 kg
2 lb / 1 kg (5)
20,000 lb / 10,000 kg
5 lb / 2 kg (4)
30,000 lb / 15,000 kg
10 lb / 5 kg (3)
50,000 lb / 25,000 kg
10 lb / 5 kg (3)
70,000 lb / 35,000 kg
20 lb / 10 kg (2)
100,000 lb / 50,000 kg
20 lb / 10 kg (2)
Note: The changes will again be saved here and the scale could be turned off if you
don't wish to change the calibration.