CS5000, Users

Rev V, January, 2008

15

2: The IZSM (initial zero setting mechanism on power up) must be within +/- 10% of the

zero obtained at calibration.

3: The push-button zero and AZSM can only operate within +/- 2% of the IZSM.
4: The overcapacity point is 103% of capacity above the IZSM.


9. The scale shows “EE-05

EE-05

EE-05

EE-05”. Press the ZERO and PRINT keys. Enter the desired

graduation. Use the following table to select a graduation value. (Press the ZERO
and PRINT keys to enter this value.)

Setting

Count by

in lb

Count by

in kg

0

100

50

1

50

20

2

20

10

3

10

5

4

5

2

5

2

1

6

1

0.5

7

0.5

0.2

8

0.2

0.1

9

0.1

0.05

10

0.05

0.02

11

0.02

0.01

NOTE: To meet stated accuracy statements, it is suggested that you choose your
graduation setting from the table below:

If your capacity is:

Set your graduation to:

500 lb / 250 kg

0.2 lb / 0.1 kg (8)

1000 lb / 250 kg

0.5 lb / 0.2 kg (7)

2,000 lb / 1,000 kg

1 lb / 0.5 kg (6)

5,000 lb / 2,500 kg

1 lb / 0.5 kg (6)

10,000 lb / 5,000 kg

2 lb / 1 kg (5)

20,000 lb / 10,000 kg

5 lb / 2 kg (4)

30,000 lb / 15,000 kg

10 lb / 5 kg (3)

50,000 lb / 25,000 kg

10 lb / 5 kg (3)

70,000 lb / 35,000 kg

20 lb / 10 kg (2)

100,000 lb / 50,000 kg

20 lb / 10 kg (2)


Note: The changes will again be saved here and the scale could be turned off if you
don't wish to change the calibration.