10. The scale displays “LL-00

LL-00

LL-00

LL-00”. Press the ZERO and PRINT keys. On the next screen

enter the capacity, in the units (lb or kg) that you are calibrating in. With no weight
on the hook, press the ZERO and PRINT keys to enter in this capacity. Max
capacity is 99999 lbs. Special versions allow a max capacity of 199999 lbs.


11. The scale shows “LL-01

LL-01

LL-01

LL-01”. Apply the first weight and press the ZERO and PRINT

keys. On the next screen enter the weight being applied. With the weight on the
hook and its exact value shown on the CS5000 screen, press the ZERO and
PRINT keys to enter in the weight.

12. The scale shows “LL-02

LL-02

LL-02

LL-02”. Apply the second weight and press the ZERO and

PRINT keys. On the next screen enter the weight being applied. With the weight
on the hook and its exact value shown on the CS5000 screen, press the ZERO and
PRINT keys to enter in the weight.


13. The scale shows “LL-03

LL-03

LL-03

LL-03”. Apply the final weight and press the ZERO and PRINT

keys. On the next screen enter the weight being applied. With the weight on the
hook and its exact value shown on the CS5000 screen, press the ZERO and
PRINT keys to enter in the weight.

Finish

The new calibration information is saved.

14. Press the calibration switch ONCE to disable calibration. This will prevent

accidental entry into the calibration mode. If you are unsure of whether the switch is
enabled or disabled, start going thorough the calibration menu again. After the
“AZT” menu item, the scale should return to normal weighing instead of saving the
AZT value and proceeding to EE-04

EE-04

EE-04

EE-04. See the “calibration switch” section.


15. Replace the access plug which will cover the calibration switch.

16. Cover this plug with a new calibration seal (e.g. foil calibration sticker).

17. Verify the calibration.

18. Calibration complete.

Legal-for-Trade Sealing

For qualified calibration facilities:
1. After calibration push the calibration switch. The calibration switch is located on the

upper right hand corner on the backside (side with battery caps) of the scale.

2. Verify that the scale returns to normal weighing when trying to access the weight

calibration.

3. Replace the set screw that covers the cal switch.
4. Apply the protection sticker (sticker must adhere to Intercomp’s specifications).