0 troubleshooting & testing, 1 test mode – Rice Lake Legal For Trade Platform/Fitness Scale - Rice Lake Dietary Fitness Scale Technical and Operating Instructions User Manual

Page 28

Advertising
background image

24

Dietary/Fitness Technical and Operating Instructions

9.0

Troubleshooting & Testing

Refer to the following instructions to check and correct any failure before contacting service personnel.

Table 9-1. Troubleshooting Table for the Rice Lake Scale Line

Symptom

Possible Cause

Corrective Action

Scale does not turn on

Dead battery

Connect the scale to a power source.

Faulty electrical outlet

Use a different electrical outlet.

Bad power supply

Replace adaptor.

Questionable weight or the scale
does not zero

External object is interfering with the
scale

Remove the interfering object from the scale.

Display did not show 0.0 before
weighing

Zero the scale and begin the weighing process
again.

Scale is not placed on a level floor

Ensure the scale is level and begin the weighing
process again.

Scale is out of calibration

Check the weight with a known weight value.

Improper tare

Place the tare item on the scale. Once the weight
of the item is displayed, press TARE.

The display shows a STOP message

The load on the scale exceeds the
capacity of the scale

Remove the excess weight and use the scale
according to manufacture’s specs.

The display shows LO Bat message

The battery is low

Recharge the battery.

The display shows Err message as
detailed below

Err 2

Low saturation state (low A/D)

The load cell is not connected properly. Check
the cables and mechanical connections. If the
problem persists, replace the set of load cells.

Err 3

High saturation state (high A/D)

See Err 2

Err 6

Unstable weight. Cannot calibrate

Check the load cells’ mechanical surroundings
and see that nothing touches them and that the
cables are properly welded.

Err 7

Mathematical error; division by zero.
Cannot calculate calibration factor.

This error will show when trying to calibrate the
unit with no calibration weight on the unit.

Calibration or configuration settings
are messed up

Pushed the wrong keys

Reset to factory defaults. See Page 9 for
instructions.

The display shows a Con En message The configuration jumper is not in

place.

Install the configuration jumper on both pins. (see
Figure 4-1 on page 7 for more information).

9.1

Test Mode

The test mode menu is a special mode used for checking three very important parameters which are useful in

knowing the system’s state and for troubleshooting. The configuration/calibration jumper must be removed

before you can enter this mode (see Figure 4-1) and power must be off.
To access the test mode parameter, press and hold the

REWEIGH

+

On-Off/Zero

keys simultaneously until middle

arrows appear in the display.
The test mode has three parameters. They are:

Weighing

Internal count

Battery indication.

Alternating between the parameters is performed by pressing the

REWEIGH

key.

Press the

Kg-Lb

key to zero the scale while displaying weight (

TEST

).