Rice Lake MSI6260cs Digital Crane Scale Software Upgrade User Manual
Page 12
Page 11 of 15
9.7 12V SLA Battery calibration.
NOTE:
This must be performed only if scale is 12V SLA battery operated. For alternative
power sources, refer to full version of MSI-9000 CellScale User’s Guide for
proper System Power settings.
9.8 Select (1) Hard-wired (Type A): AC or DC.
9.8 Select (9) Battery Calibration.
Note change to 12V SLA.
9.10 Select (1) Hard-wired (Type A): AC or DC
***** System Power *****
(1) Hard-wired (Type A): AC or DC
(5) Shutdown with: switch only
(6) Re-start with: switch only
(7) Auto Shutdown: Disabled
(8) Shutdown Now!
(9) Battery Calibration
(ESC) Exit
Current Power Level: 100 %
RAM backup battery is OK
***** System Power *****
(1) Battery (Type 1): 12V Sealed Lead Acid
(2) Check battery every: 30 seconds
(3) Send warning at: 10 % Power Level
(4) Shutdown at: 1 % Power Level
(5) Shutdown with: switch only
(6) Re-start with: switch only
(7) Auto Shutdown: Disabled
(8) Shutdown Now!
(9) Battery Calibration
(ESC) Exit
Current Power Level: 110 %
RAM backup battery is OK
***** Battery Calibration *****
Cut off voltage ADC count setting is: 3
(1) Hard-wired (Type A): AC or DC
(2) Data Backup from ROM
(Y) Save Current Calibrated Value To ROM.
(ESC) Exit
Please Set Calibration Voltage at: 9.88 V
Raw A/D counts: 187