10 data back-up – Rice Lake MSI6260cs Digital Crane Scale Software Upgrade User Manual

Page 13

Advertising
background image

Page 12 of 15

9.11 Set input voltage to 10.8VDC. Press (Y) to save value to ROM.

Note change to 12V SLA.













Note that after pressing (Y), the scale turned off. Increase input voltage to 12.5VDC and

turn scale ON.

10 Data Back-Up.


10.1 Select (1) Channels/Calibration.

















10.2 Select (2) Data backup.
















***** Battery Calibration *****

Cut off voltage ADC count setting is: 89

(1) Battery (Type 1): 12V Sealed Lead Acid

(2) Data Backup from ROM

(Y) Save Current Calibrated Value To ROM.

(ESC) Exit

Please Set Calibration Voltage at: 10.80 V

Raw A/D counts: 89

---------- MSI 9000 - CellScale ----------
- (c) 2006 Measurement Systems International -
----------------------------------------------

16:57:52 01/05/08

(1) Channels/Calibration

(2) Communications
(3) Display
(4) Feature Locks/Passwords
(5) Product IDs
(6) Set Points
(7) System
(8) About...

Enter choice :

***** Channel Setup *****

(1) Channel Calibration

(2) Data backup

(3) Filter (software)
(4) Channel Scanlist functions
(5) Motion Compensation
(6) Motion Detection
(7) Tare
(8) Total
(9) Units Lock
(a) Zero
(b) Standard Mode: Industry
(c) 4 to 20mA
(d) Temperature Sensor Installed: No
(ESC) Exit