5 Check after Installation and Start-Up

After the installation of all mounting kits and load cells, the correct mounting must be checked. In

particular, force shunts must be avoided.
After mounting also, check
• whether the load cell is inserted without tilt in the mounting kit.
• whether the upper and lower mounting plate is positioned horizontally.
• whether there are vertical space for movement and the required space for thermal expansion.
• whether there are clearance of the constrainer.
The space for movement required for displacement of the object to be weighed due to thermal

expansion, vibration, etc. can be used only if load cell is installed exactly.
To prevent vertical force shunts, all mechanical connections (pipes, cables and bellows) of the object to

be weighed to its surrounding construction must be as flexible as possible. The overall load must be

supported by the load cells.