Troubleshooting – Welch Allyn Video Colposcope - User Manual User Manual

Page 33

Advertising
background image

Directions for Use

Maintenance

31

Troubleshooting

Condition

Check

Action

Power does not come on. Power Cord

Attachment of unit to handle

Attachment of unit to handle and door
to unit

Fuses

Wall outlet

Check connections at power supply and
wall outlet.

Check colposcope unit’s alignment with
handle.

Make sure lamp access door and handle
are properly attached.

Remove fuse panel & replace blown fuse
with T1.00L 250V 1A time lag (Welch Allyn
part number # 488307-9).

Plug the power cord into a wall outlet
known to work.

No image on monitor.

Monitor/other peripheral devices

Cable connections

Monitor input selection

Make sure power is on for all devices.

Make sure all video cables are connected
properly.

Make sure correct input selection is made
on the front panel.

Lamp will not light.

Lamp housing

Lamp

Lamp access door

Make sure lamp assembly is installed
properly.

Replace lamp.

Make sure lamp access door is properly
closed.

Printer does not
respond properly to
handle buttons.

Power cord and printer

Printer

Unplug power cord from wall outlet, then
plug cord back in to outlet. At the same
time, turn printer off and then on again.

Make sure printer is on and is properly
connected.