Checking and adjusting – Leica Geosystems TPS1100 Professional Series - User Manual User Manual

Page 37

Advertising
background image

37

TPS1100 - User Manual 2.2.1en

Checking and adjusting

6

10

12

21

Line of sight, continued

Confirms readiness for
measurement and goes to

the measuring menu.

Sight the target accurately.

Perform the second
measurement.

Stores the new values.
Repeats the complete line-of-
sight error determination
procedure.
Leaves the old values
unchanged.

If the value of the line-of-
sight error (c) exceeds 5' 24"

(0.1 gon), the collimation error is to
be redetermined. If the limit is
repeatedly exceeded, please notify
service.

When the new line-of-sight error has
been confirmed, the tilting-axis error
can be determined.

Confirms that the tilting-axis
error should also be

determined.

Ends the function and returns
to the calibration dialog.

After successful completion of the
second measurement, the older and
the newly-determined line-of-sight
errors are displayed.

Main\ Hz-Collim. Error

old new

c Hz-c: -0°00'03" 0°00'08"

Accept new value(s)?

YES RETRY

NO

MC

QUIT

Main\ Hz-Collim. Error

YES

NO

MC

Continue to determine the

tilting-axis errror?

Telescope Positioning

Hz- and V-positioning:

set direction(s) to zero.

∆∆∆∆∆

Hz

:

0°00'00"

∆∆∆∆∆

V

:

0°00'00"

OK

ABORT

MC

QUIT

Main\ Hz-collim. Error

Aim accurately at the same

target in other face

Hz

:

193°19'24"

V

:

269°08'45"

MEAS

MC

QUIT