3 Technical data

13

Melting Point M-560 Operation Manual, Version D

Table 3-4: Technical data of the Melting Point M-560

Melting Point M-560

250 °C to 400 °C

±0.3 °C to ±0.5 °C

Repeatability of melting point at 0.5 °C/min

±0.1 °C

Accuracy of boiling point at 1.0 °C/min up to 400 °C

±0.5 °C

Repeatability of boiling point at 1.0 °C/min

±0.3 °C

Temperature gradients, °C/min

0.1, 0.2, 0.5, 1, 2, 3, 5, 10, 20

Heat-up time (50 °C–350 °C) at 25 °C

~4 min

Cool-down time (350 °C–50 °C) at 25 °C

~13 min

Electrical supply

100–240 V (±10 %), 50–60 Hz

Power consumption

150 W

Contact termination

L, N, PE

Approval

CE, CSA, UL

Dimensions (W×H×D), mm

190Ч200Ч370

Weight, kg

4.5

Environmental conditions

For indoor use only

Temperature

5–40 °C

Altitude

up to 2000 m a.s.l.

Humidity

Maximum relative humidity 80 % for temperatures up to
31 °C, decreasing linearly to 50 % relative humidity at 40 °C

Overvoltage category

II

Degree of protection

IP20

Pollution degree

2

Storable methods for melting point

50

Storable methods for boiling point

50

Compliant with Pharmacopeia methods

PH. EUR. 6.1.2.2.60, USP XXI 741 and JP

NOTE
Temperature measuring accuracy refers to pharmacopoeia melting point.

3.3

Materials used

Table 3-5: Materials used

Component

Material designation

Print holder

PA

Heating block

Aluminium

Lenses

Glass

Axial fan

Aluminium

Housing

PU, Stainless Steel, Glass

Cover

POM, Ceramic, Aluminium, Stainless Steel