BUCHI Melting Point M-560 User Manual

Page 23

Advertising
background image

6 Operation

23

Melting Point M-560 Operation Manual, Version D

6.2.2

Determination without pre-registered melting point method

• Choose the menu for melting point determination.
• Make sure that all samples are removed.
• Press START to begin the determination process immediately with the last parameters used. They

always remain saved in the instrument and are indicated in the green field in the lower part of the
screen.

NOTE
If the current parameters do not meet your requirements, press Edit and set start temperature, stop
temperature, and temperature gradient.

a b

c

d

a Start temperature
b Stop temperature
c Temperature gradient
d Edit

• To adjust the desired melting parameters,

press

Edit.

• Chose

start temperature

with the rotary knob, and
press

Next.

• Chose

stop temperature

with the rotary knob, and
press

Next.

• Chose

temperature

gradient with the rotary
knob, and press

Save.

NOTE
Insert samples only when the start temperature is reached (as soon as prompted).