6 Operation

35

Melting Point M-560 Operation Manual, Version D

6.4.4

Verification

To verify the calibration, proceed as follows:
• Measure all standards (BUCHI recommends using the verification set (11055019) but it is also

possible to use your internal standards.

NOTE
The verification is not a guided process.

• When all measuring results were within the required tolerances, click on

Verified. The following

screen appears:

• Enter the following password:

VER. The current date is now indicated under “Last verification”.

6.5

Settings, SysInfo, Test

Test: Functional test of the instrument. To
perform the test of several functions follow
carefully the on-screen instuctions. If a function
test does not show

ok, please contact the cus-

tomer service. The Test protocol is described in
section 6.5.1.
SysInfo: Instrument gives information regarding
settings and connected devices. The SysInfo
protocol is described in section 6.5.2.
Settings: This menu can be used to change
parameters.